19mm (0.75“) Digit Height
Cold Junction Compensation
Simple Connection
LED Backlighting
DTM 995 LCD Type K Thermocouple Indicator
CALIBRATION
1. Place thermocouple at 0°C and adjust OFFSET to
read 000.
2. Place thermocouple at known reference
temperature, eg. 100°C and adjust CAL
to read 100.
3. Repeat steps 1 and 2 until there are no more
adjustments of CAL and OFFSET required to
give desired readings.
In room temperature application adjust CAL to give the
same reading as a known temperature reference.
This meter will give a direct reading in °C from a K type (chromel-alumel) thermocouple sensor. The circuit uses a converter/amplifier
which has Cold Junction Compensation. Calibration is by two 20-turn cermet potentiometers which allow sensitive adjustment of the
instrument. The meter can also operate without a thermocouple to read room temperature. The meter is housed in a robust carrier,
which can be bolted in place or panel mounted using the low profile bezel, window and clips provided. The meter can also be backlit for
use in poor visibility.
Specification Min. Typ. Max. Unit
Accuracy (overall error) *** 1 % (±1 count)
Thermocouple measurement range
-50 +600 °C
-58 +1112 °F
Power supply voltage (V+ to V-)** 7.5 9.0 24 V
Operating temperature range * 0 50 °C
Temperature stability 0.05 °C/°C
Supply current 3.5 mA
Supply current with backlighting 55 mA
Stock Number
Standard Meter DTM 995
CONNECTOR SOURCING GUIDE
Screw Terminals - No Connector RequiredMETHOD
ON-BOARD SOLDER LINKS
* Module temperature
**
*** To ensure maximum accuracy, re-calibrate periodically
5V supply option available - Contact Sales for details
REAR
VIEW
T- (blue)
T+ (brown)
V-
LP-
V+
LP+
CAL
OFFSET
61.5 (2.42)
51.0 (2.01)
Ø5.7
(0.22)
24.5 (0.96)
24.0 (0.94)
39.0 (1.54)
Ø3.0
(0.12)
71.5 (2.81)
d
e
76.0 (2.99)
72.5 (2.85)
40.5 (1.59)
44.0 (1.73)
24.0 (0.94)
51.0 (2.01)
a
b
c
Panel cut-out 72 x 40 (2.83 x 1.57)
Maximum panel thickness 3mm (0.12).
a. 1.0 (0.04)
b. 2.0 (0.08)
c. 6.5 (0.26)
d. 11.5 (0.45)
e. 13.0 (0.51)
2
3
SAFETY
The user must ensure that the incorporation of the DPM into the user’s equipment conforms to the relevant sections of BS EN 61010 (Safety Requirements
for Electrical Equipment for Measuring, Control and Laboratory Use).
www.lascarelectronics.com
Vue de la page 0
1 2

Résumé du contenu

Page 1 - CONNECTOR SOURCING GUIDE

DIMENSIONSAll dimensions in mm (inches)19mm (0.75“) Digit HeightCold Junction CompensationSimple ConnectionLED BacklightingDTM 995 LCD Type K Thermoco

Page 2 - PANEL FITTING

PIN FUNCTIONSROOM TEMPERATURE METER1. V+ Positive power supply connection.2. V- Negative power supply connection.3. T+ Connection for positive (Ni-CR)

Commentaires sur ces manuels

Pas de commentaire