Martel-electronics PSC4010 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Martel-electronics PSC4010. Martel Electronics PSC4010 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Reference Manual

PSC4010Reference Manual

Page 2

8[DONE]. The [FREQ LEVEL] option is used to adjust the amplitude of the wave. [NEXT] isused to access the contrast main menu, and [DONE] returns to th

Page 3 - 1. Introduction

92.3 Cursor control / Setpoint controlThe output value can be controlled by the four cursor control arrows on the keypad. Bypressing one of the arrows

Page 4 - 1.3 Safety information

103. Using Measure Modes3.1 Measuring volts, millivolts and frequencyThe calibrator can measure volts or frequency on these jacks. To make the desired

Page 5

113.3 Measuring mA with Loop PowerTo test a 2-wire, loop powered transmitter that is disconnected from wiring, use the looppower function. This functi

Page 6 - 2. Calibrator Interface

124.2 Using the Automatic Output FunctionsThere are two automatic output functions, step and ramp. The selected function can beturned on and off using

Page 7

134.3-1 HART™ Resistor SelectionThe PSC4010 can be set-up so that the 250 ohm resistor required for Hart™ configurationdevices resides inside the PSC

Page 8 - 2.2 Menu Bar

144.5 Sourcing voltsTo source volts follow these steps:1. Select the [CONFIG] option from the main menu.2. Choose [VOLTS] from the primary parameters.

Page 9 - CONFIG MORE

154.7 Sourcing frequencyTo source a signal use these steps:1. Select the [CONFIG] option from the main menu and select frequency from the primaryparam

Page 10

165. Remote OperationThe calibrator can be remotely controlled using a PC terminal, or by a computer programrunning the calibrator in an automated sys

Page 11

176. Select Ok7. To see if the port works enter *IDN?. This command will return information on the unit.5.2 Changing Between Remote and Local Operatio

Page 12 - 0.000 mA

PSC4010 Reference Manual1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 13 - 4. Using Source Modes

18Overlapped CommandsCommands that require more time to execute than normal. The command *WAI can be usedafter the overlapped command to tell the cali

Page 14

195.3-4 Calibrator StatusStatus registers, enable registers, and queues provide status information on the calibrator.Each status register and queue ha

Page 15 - 4.4 Simulating a Transmitter

207 6543 210PON 0 CME EXE DDE QYE 0 OPCPONPower On. Set to 1 if power was turned on and off before the Event Status Registerwas read.CMECommand Error.

Page 16 - 4.000 mA

215.4 Remote Commands and Error CodesThe following tables list all commands, and their descriptions, that are accepted by thecalibrator. Table 5: Comm

Page 17 - 1.0 kHz

22Command DescriptionPULSE_CNT? Returns the number of pulses in the pulse trainREMOTE Puts the calibrator in remote modeSIM Sets the output for mA sim

Page 18 - 5. Remote Operation

5.5 Entering CommandsCommands for the calibrator may be entered in upper or lower case. There is at least onespace required between the command and p

Page 19 - 5.3 Using Commands

24*OPC?Returns 1 when all operations are complete, and causes program execution to pause until allthe operations are complete. For example:TRIG ; *OPC

Page 20

25For example, if a value for current output is entered that is bigger than the supported range(0-24mA) FAULT? Would return:103 which is the code numb

Page 21

26LOCALRestores the calibrator to local operation if it was in remote mode. Also clears LOCKOUT if theunit was in lockout mode.MEASSets the display to

Page 22

27SIMSets the output for current simulation. This command also switches the calibrator into mAsimulation mode. A number and a unit must be entered aft

Page 23 - Table 6: Calibrator Commands

11. IntroductionThe Martel PSC4010 Multifunction Process Calibrator is a handheld, battery-operatedinstrument that measures and sources electrical and

Page 24 - Table 8: Error codes

286. SpecificationsOne year at 23°C ± 5°C, unless specified otherwise. Outside of this range the stability of themeasurements is ± 0.003% of FS/°C.Tab

Page 25 - 5.5 Entering Commands

Table 13: Frequency Measurement/SourceRange Accuracy (% of reading ± floor)Read 2.0CPM - 600.0CPM 0.05% ± 0.1CPM1.0Hz - 1000.0Hz 0.05% ± 0.1Hz1.00KHz

Page 26

7. Maintenance / Warranty7.1 Replacing BatteriesReplace batteries as soon as the battery indicator turns on to avoid false measurements. Ifthe batteri

Page 30 - 6. Specifications

www.martelcorp.come-mail: [email protected]: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Martel Electronics3 Corporate Park Dr.Derry

Page 31

1.3 Safety information Symbols UsedThe following table lists the International Electrical Symbols. Some or all of these symbolsmay be used on the inst

Page 32 - 7. Maintenance / Warranty

WarningTo avoid possible electric shock or personal injury:• Do not apply more than the rated voltage. See specifications for supported ranges. • Foll

Page 33

2. Calibrator InterfaceFigure 1 shows the location of the input and output connections on the calibrator, while Table1 describes their use.Table 1: I

Page 34

5Figure 2 shows the location of the keys on the calibrator. Table 2 lists the functions of eachkey.Table 2. Key FunctionsNo. Name Function1 Function K

Page 35

6The display of the calibrator is shown in Figure 3. The menu bar is used to setup the display to perform the desired function.Table 3 explains the di

Page 36 - 0219895 Rev C 5/11

7The pulse home menu also has three active options, [MENU], [TRIG], and [COUNTS]. The[TRIG] and [COUNTS] options are used for pulse simulation. The fu

Commentaires sur ces manuels

Pas de commentaire