Martel-electronics MC1210 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Martel-electronics MC1210. Martel Electronics MC1210 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Reference Manual

MC1210Reference Manual

Page 2

101. Turn on the unit while holding down the “HOME” key.2. Hold the key down for a count of 10 seconds to restore contrast default settings.If the dis

Page 3 - 1. Introduction

112.3 Cursor control / Setpoint controlThe output value can be controlled by the four cursor control arrows on the keypad. Bypressing one of the arrow

Page 4 - 1.3 Safety information

123. Using Measure Modes (Lower Display)3.1 Measuring volts and frequencyElectrical parameters volts and frequency can be measured using the lower dis

Page 5

3.3 Measuring Temperature3.3-1 Using ThermocouplesThe calibrator supports the followingthermocouple types: B, C, E, J, K, L, N, R, S, T,U, BP, and XK.

Page 6 - 2. Calibrator Interface

3. Choose 2, 3, or 4-wire connection [2W, 3W, 4W]. (4-wire allows for the most precisemeasurement)4. Select RTD type from the sensor types. 5. Attach

Page 7 - 2.1 Main Display

15To measure pressure, follow these steps:1. Connect the pressure module to the calibrator as shown in Figure 9.The calibrator can measure pressure on

Page 8 - 2.2 Menu Bar

164. Using Source Modes (Lower Display)The calibrator can generate calibrated signals for testing and calibrating process instruments.It can source vo

Page 9

17Figure 10. Connections for Sourcing Current4.3-1 HART™ Resistor SelectionThe MC1210 can be set-up so that the 250 ohm resistor required for Hart™ c

Page 10

184.4 Simulating a TransmitterTo have the calibrator supply a variable test current to a loop in place of a transmitter, followthese steps:1. Select l

Page 11

194.6 Sourcing frequencyTo source a signal use these steps:1. Switch to the lower display and select frequency from the primary parameters.2. Select o

Page 12 - 3.2 Measuring mA

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31.1 Cu

Page 13 - 3.3 Measuring Temperature

20Figure 13. Connections for Outputting Thermocouples4.8 Sourcing ThermocouplesTo source a thermocouple use the following steps:1. Connect the thermoc

Page 14 - Warning!

4.9 Sourcing Ohms/RTDsTo source an RTD, follow these steps:1. Select lower display from the Main Menu, and choose [RTD] from the primaryparameters.2.

Page 15 - PRESSURE

22temperature, maximum temperature, R0, and the values for each of thetemperature coefficients. The custom function uses the Calendar-Van Dusen equati

Page 16 - 4.3 Sourcing mA

23Figure 17. Connection Using Current Loop5.2-1 HART™ Resistor SelectionThe MC1210 can be set-up so that the 250 ohm resistor required for Hart™ conf

Page 17

24CautionTo avoid mechanically damaging the pressure module, never apply more than 10 ft-lb. oftorque between the pressure module fittings, or betwee

Page 18 - 4.5 Sourcing volts

6. Using the Upper and the Lower Display for Calibration andTesting6.1 Testing an Input or Indicating DeviceTo test and calibrate actuators, recording

Page 19 - 4.6 Sourcing frequency

266.3 Calibrating a Transmitter To calibrate a transmitter both the upper and the lower displays will be used; one formeasuring and the second a sourc

Page 20 - 4.8 Sourcing Thermocouples

27Figure 22. Calibrating a Pressure Transmitter7. Performing a Switch TestThe calibrator can stimulate a switch and capture the results in any source/

Page 21 - 4.9 Sourcing Ohms/RTDs

281. Use the menu to select SW TEST on the upper display.2. Connect the calibrator to the switch using the switch terminals. The polarity of the termi

Page 22 - 5.1 Measuring volts and mA

7.2 Performing a pressure switch test:Figure 24. Pressure Switch Test1. Use the menu to select SW TEST on the upper display.2. Connect the calibrator

Page 23

31. IntroductionThe Martel MC1210 Multifunction Process Calibrator is a handheld, battery-operatedinstrument that measures and sources electrical and

Page 24

307. At the top of the display it will now read, “SW OPENED AT” and give you the pressure thatthe switch opened at.8. Press the “NEXT” option to view

Page 25 - 6.2 Calibrating an I/P Device

318. Remote OperationThe calibrator can be remotely controlled using a PC terminal, or by a computer programrunning the calibrator in an automated sys

Page 26

326. Select Ok7. To see if the port works enter *IDN?. This command will return information on the unit.8.2 Changing Between Remote and Local Operatio

Page 27 - 7. Performing a Switch Test

Compound CommandsCommands that contain more than one command on one line. For example:LOWER_MEAS RTD; RTD_TYPE CU10Sets the calibrator to measure RTD

Page 28

34Indefinite ASCII (IAD)Any ASCII characters followed by a terminator. For example:*IDN? returns MARTEL, MC1210, 250, 1.008.3-4 Calibrator StatusStatu

Page 29 - SW TEST CLOSE

ESR, and setting it to 0 disables the corresponding bit. Cleared at power reset. Bitassignments for the ESR and the ESE respectively are shown below.1

Page 30

36Input BufferCalibrator stores all received data in the input buffer. The buffer holds 250 characters. Thecharacters are processed on a first in, fir

Page 31 - 8. Remote Operation

Command DescriptionCPRT_COEFC Sets the custom RTD coefficient CCPRT_COEFC? Returns the custom RTD coefficient CCPRT_MIN_T Sets the custom RTD minimum

Page 32 - 8.3 Using Commands

Command DescriptionTC_TYPE? Returns the thermocouple typeTEMP_UNIT Sets input/output temperature unit for RTD and TCTEMP_UNIT? Returns the temperature

Page 33

Units MeaningMMH2O4C Pressure in millimeters of water at 4°CMMH2O20C Pressure in millimeters of water at 20°CFTH2O60F Pressure in feet of water at 60°

Page 34

41.3 Safety information Symbols UsedThe following table lists the International Electrical Symbols. Some or all of these symbolsmay be used on the ins

Page 35

*ESELoads a byte into the Event Status Enable register. The command is entered with a decimalnumber that, when converted to binary, enables the right

Page 36

41*SRE?Returns a byte from the SRE. The byte is returned in decimal format. For example:If 40 is returned, bits 5 and 3 are enabled.*STBReturns the st

Page 37

42CPRT_COEFA?Returns the number which was entered for the first coefficient of the polynomial used in thecustom RTD. Using the example above CPRT_COEF

Page 38 - Table 7: Parameter units

43CPRT_MAX_TSets the maximum temperature of the custom RTD range. The temperature value must beentered with a degrees label, CEL for Celsius and FAR f

Page 39 - 8.5 Entering Commands

FREQ_TYPEWhen in frequency mode, sets the calibrator to output a continuous wave (Frequency Out), ora pulse train. To set the calibrator to continuous

Page 40

L_PRES_UNIT KPAL sets the pressure unit to kiloPascalsOUTSets the output of the calibrator. This command may be used to output mA, volts, frequency,t

Page 41

46RTD_TYPESets the RTD type. The following is a list of RTD types they way they should be entered afterthe command:PT385_10; PT385_50; PT385_100; PT38

Page 42

47ST_STARTStarts a switch test.TC_TYPESets the type of the thermocouple. All available types are shown in the TC Types table inSection 8. (Specificati

Page 43

U_PRES_UNITSets the unit for measuring pressure on the upper display. Add the unit after the command.The available pressure units and their syntax are

Page 44

499. SpecificationsOne year at 23°C ± 5°C, unless specified otherwise. Outside of this range the stability of themeasurements is ± 0.003% of FS/°C.Tab

Page 45

WarningTo avoid possible electric shock or personal injury:• Do not apply more than the rated voltage. See specifications for supported ranges. • Foll

Page 46

50Table 12: Frequency Measurement/SourceRange Accuracy (% of reading ± floor)Read 2.0CPM - 600.0CPM 0.05% ± 0.1CPM1.0Hz - 1000.0Hz 0.05% ± 0.1Hz1.00KH

Page 47

51T -250.0 -200.0 1.5 1.7-200.0 0.0 0.5 0.70.0 400.0 0.2 0.4E -250.0 -200.0 1.0 1.2-200.0 -100.0 0.3 0.5-100.0 1000.0 0.2 0.4R10.0 200.0 1.7 1.9200.0

Page 48

RTD Type Range (°C) Accuracy (°C)Minimum Maximum Measure 4W/Source Measure 3W300.0 400.0 0.5 0.7400.0 630.0 0.5 0.8630.0 800.0 0.6 0.9PT385, 100

Page 49 - 9. Specifications

53RTD Type Range (°C) Accuracy (°C)Minimum Maximum Measure 4W/Source Measure 3WPT385, 1000 ohm -200.0 -80.0 0.2 0.2-80.0 0.0 0.2 0.20.0 100.0 0.2

Page 50

5410.3 Service Center Calibration or Repair Only qualified service personnel should perform calibration, repairs, or servicing not coveredin this manu

Page 52

www.martelcorp.come-mail: [email protected]: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Martel Electronics3 Corporate Park Dr.Derry

Page 53 - 10. Maintenance / Warranty

2. Calibrator InterfaceFigure 1 shows the location of the input and output connections on the calibrator, while Table1 describes their use.Table 1: I

Page 54 - 10.5 Warranty

7Figure 2 shows the location of the keys on the calibrator. Table 2 lists the functions of eachkey.Table 2. Key FunctionsNo. Name Function1 Function K

Page 55

The display of the calibrator, shown in Figure 3, is divided into three main sections: the upperdisplay, the lower display, and the menu bar. The upp

Page 56 - 0220001 Rev F 3/13

9In the input home menu the only active options are [MENU] and [LIGHT]. The [MENU] optionis used to enter the next level of the menu bar, the main men

Commentaires sur ces manuels

Pas de commentaire