BetaGauge PIR/PIR PROReference Class High Accuracy DigitalPressure Test GaugeUser Reference ManualThis documentincludes informationfor both the CSA an
RS-232 InterfaceAn RS-232 interface is standard on the BetaGauge PIR. Serialcommunication can be used for configuration, calibration, and totransfer
Standard Engineering UnitsSee Table of Ranges and Resolutions, page 13, plusOne custom unit (user programmable)Media CompatibilityAll ranges except t
Appendix 1: BetaGauge PPIIRRCalibration ProcedureOverviewCalibration adjustment of the BetaGauge PIR is performedelectronically via internal software
Procedure:Screens shown in this manual represent the displays shown with a500 psi Gauge. The BetaGauge PIR will prompt the technician forthe appropri
unit takes readings, the screen will show ———-. When thereadings are complete the screen should look as shown below.Use the Pressure Standard to outp
Use the Pressure Standard to output 0.00 psi. After the output hasstabilized, press the ENTER key to continue. As the unit takesreadings, the screen
14BetaGauge PI Reference Class Ranges and ResolutionsPressure range 30 100 300 500 1000 3000 5000 10000gauge ** gauge **Range Type compound gauge ** g
9. BetaGauge PIR Serial InterfaceInstructionsInitiating CommunicationThe terminal communications can be setup using terminalcommunication software on
Parameter UnitsUnits MeaningPsi Pressure in pounds per square-inchBar Pressure in bars mBar Pressure in millibarsKg/cm2 Pressure in kilograms per cent
Error CodesError Description101 A non-numeric entry was received where it should be anumeric entry102 Too many significant digits entered103 Invalid u
10. WarrantyMartel Electronics Corporation warrants all products againstmaterial defects and workmanship for a period of twelve (12)months after the d
19
20
21
P/N 0200018 Rev. G 5/13www.martelcalibrators.come-mail: [email protected]: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Ma
1. IntroductionThe BetaGauge PI Reference Class (PIR) combines the highaccuracy of digital electronics with the convenience and ease ofuse of an analo
Symbol DescriptionPower OFFPower ONEarth groundRisk of Danger. Important information. Refer to manual.BatteryHazardous VoltageConforms to ATEX require
Special Conditions for Safe Use:MisuseShould the BetaGauge PIR be exposed to overpressure orsudden physical shock (i.e. being dropped) it should be ex
3. OperationPower: The standard BetaGauge PIR is supplied with 3 AAbatteries installed. If you purchased the optional 24Volt poweredversion, batteri
6. Set sample rate: This determines how often pressure issampled and the display is updated. Choices are 0.5, 1, 3,and 10 samples/second. Note that
• Auto shutdown adjustment (enable/disable)• Damping settings (enable/disable)• Sample rate setting (enable/disable)• Tare setting (enable/disable
gauge, the displayed pressure will not reflect the true pressureapplied.6. Changing the BatteriesWARNINGExplosion hazardBatteries must only be changed