BetaGauge PI & PI PRODigital Pressure Test GaugeUser Reference ManualThis documentincludes informationfor both the CSA andCSA/ATEX certified versi
88. Specifications Available Input RangesSee page 13 for a table of available ranges in PSI plus equivalentranges and resolution for all engineering u
MechanicalDimensions 4.5” (diameter) x 2.2” (depth) x 5” (height)(11.4cm x 5.6cm x 12.7cm) (PI PRO depth is 1.5” or 3.7cm)Pressure Connection: ¼” NPT
10vacuum to the full-scale range of the unit under test. In order tomaintain the specified accuracy of the BetaGauge PI, standards shouldhave a TUR o
Use the Pressure Standard to output 375.00 psi (75%). After theoutput has stabilized, press the ENTER key to continue. As the unittakes readings, th
12Use the Pressure Standard to output 125.00 psi (25%). After theoutput has stabilized, press the ENTER key to continue. As the unittakes readings,
Pressure range 0.4 1 15 30 15 30 100 300 500 1000 2000 3000 5000 10000gauge gauge gauge ** gauge **Range Type compound compound compound compound abso
149. BetaGauge PI Serial Interface InstructionsInitiating CommunicationThe terminal communications can be setup using terminalcommunication software o
15CAL_STORE Stores calibration data.HC_AUTO_OFF Turns auto shutdown offHC_AUTO_ON Turns auto shutdown onCUST_MULT? Sets the multiplier for the custom
16Error CodesError Description101 A non-numeric entry was received where it should be anumeric entry102 Too many significant digits entered103 Invalid
10. WarrantyMartel Electronics Corporation warrants all products against materialdefects and workmanship for a period of twelve (12) months after thed
P/N 0219178 Rev M 4/13 Relatedwww.martelcorp.come-mail: [email protected]: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Martel El
11. IntroductionThe BetaGauge PI combines the high accuracy of digital electronicswith the convenience and ease of use of an analog test gauge.Accurat
2Symbols UsedThe following table lists the International Electrical Symbols. Some or allof these symbols may be used on the instrument or in this manu
Special Conditions for Safe Use:MisuseShould the BetaGauge PI be exposed to overpressure or suddenphysical shock (i.e. being dropped) it should be exa
4batteries are not installed. Connect a 24V power supply to the terminalblock on the rear of the gauge, noting proper polarity. Do not install batter
The tare value is set manually with the ▲ and ▼ keys, and isbased on the engineering units and resolution selected for display.TARE value can be set t
65. Setting a custom engineering unit or scale: The last menu choice insupervisory mode is SET FACTR. This allows you to set a multiplierfactor from
clip. Reassemble the case making certain that the face is properlyoriented.If you purchased the optional 24 Volt powered version, the terminals forpow